Vergleich einer nicht-invasiven Messung von hämodynamischen Parametern mit dem NICaS™ CS (Non-Invasive Cardiac System) mit invasiv gemessenen Rechtsherzkatheterdaten bei Patienten mit Herzleistungsschwäche
Loading...
Files
Date
Authors
Publisher
Abstract
Summary of the Study
Introduction and Background of the Study: Heart failure is a complex clinical syndrome resulting from structural and/or functional cardiac abnormalities. It typically manifests with symptoms such as dyspnea, peripheral edema, and fatigue, and is often accompanied by clinical signs such as elevated jugular venous pressure, pulmonary crackles, and lower leg edema. The underlying myocardial dysfunction results in elevated intracardiac filling pressures and/or inadequate cardiac output at rest or during exertion. With an estimated global prevalence of approximately 64.3 million affected individuals, heart failure represents a significant health and socioeconomic burden. In industrialized countries, the prevalence among adults is approximately 1–2 %. In this context, hemodynamic characterization plays a central role in diagnostics, therapeutic monitoring, and disease follow-up. The gold standard for hemodynamic assessment remains right heart catheterization (RHC), which provides comprehensive hemodynamic assessment and monitoring through direct measurement of right-sided intracardiac and pulmonary artery pressures, central mixed venous oxygen saturation and cardiac output. However, due to its invasiveness, potential complications, and the associated personnel and technical demands, this method is of limited suitability for repeated use, particularly in outpatient settings. A promising non-invasive alternative is regional impedance cardiography, which enables both intermittent and continuous measurement of hemodynamic parameters such as cardiac output, cardiac index and total peripheral resistance, along with other parameters, with relatively lower training demands and minimal risk. Its simplicity and reliability in repeated applications make this method especially suitable for routine clinical practice. However, large-scale studies systematically evaluating the diagnostic accuracy and agreement of this method compared to the reference methods are still lacking. The aim of this study was therefore to compare the hemodynamic parameters obtained non-invasively using the NICaS™ CS (Non-Invasive Cardiac System) and those measured invasively via right heart catheterization. The analysis was conducted in a real-world clinical setting in a cohort of patients with heart failure at our clinic.
Study Population and Methods: A total of 203 patients with chronic heart failure who underwent both right heart catheterization and regional impedance cardiography using the NICaS™ CS (Non-Invasive Cardiac System, NI Medical, Israel) between March 2021 and December 2023 at RHÖN-KLINIKUM Campus Bad Neustadt were enrolled in the study. Patients with severe aortic or mitral valve regurgitation, severe aortic valve stenosis, a body mass index > 40 kg/m², severe peripheral edema, advanced peripheral arterial occlusive disease, psychomotor agitation, and Arrhythmia absoluta with a heart rate > 130 bpm at the time of measurement were excluded from the study. Invasive hemodynamic measurements were performed using a Swan–Ganz catheter introduced via the femoral vein. Cardiac output was calculated using the Fick principle. The RIC measurement with the NICaS™ CS was performed either the day before, on the same day, or the day after the right heart catheterization procedure. Both assessments were performed independently and in a blinded manner. For the calculation of non-invasively determined pulmonary vascular resistance, the systolic pulmonary artery pressure was obtained by echocardiography, and a constant pulmonary capillary wedge pressure of 20 mmHg was assumed in all cases.
Results: The statistical analysis revealed no significant differences between non-invasively measured cardiac output (NI-CO: 4.1 ± 1.3 L/min) and the invasively measured reference value (RHC-CO: 4.0 ± 1.2 L/min; p = 0.1). The same applied to the cardiac index (NI-CI and RHC-CI: 2.1 ± 0.6 L/min/m²; p = 0.06). In contrast, significant differences were observed in measurements of total peripheral resistance (TPR) and pulmonary vascular resistance (PVR). Parameters obtained by regional impedance cardiography were markedly higher than the respective invasive reference parameters. Specifically, non-invasive TPR was 2225 ± 3314 dyn·s·cm⁻⁵ compared to invasive TPR of 1733 ± 590 dyn·s·cm⁻⁵ (p = 0.001), and non-invasive PVR was 386 ± 218 dyn·s·cm⁻⁵ compared to invasive PVR of 220 ± 162 dyn·s·cm⁻⁵ (p < 0.001), indicating a systematic overestimation of resistance parameters by RIC. The Pearson correlation analysis revealed significant correlations between the two measurement modalities. The strongest correlation was observed for cardiac output (r = 0.47; p < 0.001), followed by cardiac index (r = 0.38; p < 0.001) and PVR (r = 0.36; p < 0.001). For TPR, a statistically significant but weak correlation was found (r = 0.21; p = 0.02). Analysis of systematic deviations showed a bias and precision of –0.14 ± 2.56 L/min for cardiac output and –0.09 ± 1.3 L/min/m² for cardiac index. Systematic differences were more evident in the resistance parameters (TPR: –1243 ± 3510 dyn·s·cm⁻⁵; PVR: –121 ± 504 dyn·s·cm⁻⁵). A supplementary comparison of echocardiographically estimated versus invasively measured systolic pulmonary artery pressure revealed a good correlation (r = 0.71; p < 0.001). The correlations remained almost unchanged after excluding patients with severe tricuspid regurgitation.
Conclusion: Our study demonstrates that regional impedance cardiography using the NICaS™ CS (Non-Invasive Cardiac System) – with careful methodological application and consideration of predefined inclusion and exclusion criteria – can be considered a clinically feasible, non-invasive method for assessing cardiac output and cardiac index in patients with heart failure. The non-invasively measured resistance parameters – total peripheral resistance (TPR) and pulmonary vascular resistance (PVR) – showed statistically significant but overall weak correlations with their respective invasive reference values. Further prospective, multicenter studies are needed to validate these correlations.